Sign In | Join Free | My howtoaddlikebutton.com
China China World Technology Medical Equipment Service Group logo
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
Verified Supplier

5 Years

Home > Patient Monitor Parts >

PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8

Brand Name : PHiliph

Model Number : M8048A FMS-8

MOQ : 1pcs

Payment Terms : Full before shipment

Delivery Time : 3-5 working days

Packaging Details : Carton

Product Name : M8048A FMS-8 PHiliph MP60 MP70 Patient Bedside Monitor Module Rack

Brand : PHiliph

Part no. : M8048A FMS-8

Weight : <3500g ,<7.7 lb

Weight*Height *Width : <315mm*130mm*120mm (12.4*5.1*4.7inch )

Warranty : 90 days

Condition : New & Original

MOQ : One piece

Payment Method : T/T ,Paypal ,Western Union

Color : White

Contact Now

M8048A FMS-8 PHiliphhilip MP60 MP70 Patient Bedside Monitor Module Rack

PHiliphhilip FMS module rack

MAX weight:<3500g ,<7.7 lb

Weight*Height *Width:<315mm*130mm*120mm (12.4*5.1*4.7inch )

Flexible Module Server Troubleshooting

Symptoms Cause of Failure Failure Isolation and Remedy
Prompt Message “Unrecognized Measurement Module in slot s” is issued An unsupported module has been plugged into the Flexible Module Server Unplug the unsupported module
Prompt message “Measurement Module in slot n is currently ignored” is issued Too many modules of the same kind have been plugged into the Flexible Module Server Unplug module in slot n
Red Error LED stays on Unrecoverable hardware selftest error: Try connecting the MMS directly to the MSL cable. If the measured value is displayed on the screen, the fault is in the FMS
MSL cable defective If no measurements are displayed when the MMS is directly connected to the MSL cable, replace the MSL cable
Flex connector from main board to MSL defective Replace connector
CPU module defective Replace CPU module
Red Error LED flashes Hardware selftest error If system comes up, check status log. Otherwise see above
Flexible Module Server LEDs OK, Front Measurement Module not recognized (silent or INOP) Damaged measurement module or measurement module connector Replace Measurement Module
No Front-End power because MSL voltage from the monitor is too high or too low

1. Try a new MSL cable. Replace if failure is rectified.

2. Replace mother board

Mother board or connector on Mother Board defective If the voltage is within range, or if the connector is visibly damaged, replace the motherboard

PHiliphhilip FMS/MMS Module Lock

The following parts and hardware are included in this mounting kit

Item # Description Qty
1 Main Enclosure 1
2 Support Bracket 1
3 #8-32 x 1/4'' Button Head Machine Screw, Tamper-proof 4
4 5/32'' Hex Wrench, Tamper-proof 1

Mounting Notes: The bracket is designed to accommodate FMS (handle on both ends) or MMS extension (handle on one end). Installation is only shown on FMS with MMS extension.

Installing the Module Lock

1. After installing the required modules, slide the FMS into the main housing as shown below. Installation Note: It is not necessary to remove the previously installed mounting clips from the FMS to install the module lock.

2.Slide the support bracket over the MMS extension and into the main housing as far as it will go. Note the slot on the bracket for mounting the module lock on the second handle.

3.Using the 5/32'' tamper-resistant hex key provided, install four (4) #8-32 x 1/4'' tamper-resistant screws as shown below.

Installation Notes: Module locks are designed to provide space for individual modules to be disconnected without removing them from the server.


Product Tags:

MP70 Patient Monitor Module Rack

      

M8048A Patient Monitor Module Rack

      
Wholesale PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8 from china suppliers

PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8 Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: China World Technology Medical Equipment Service Group
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Inquiry Cart 0